Prevalence of Cancelled Shock Therapy and Relationship to Shock Delivery in Recipients of Implantable Cardioverter-Defibrillators Assessed by Remote Monitoring

期刊

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
卷 32, 期 -, 页码 S42-S46

出版社

WILEY-BLACKWELL PUBLISHING, INC
DOI: 10.1111/j.1540-8159.2008.02288.x

关键词

implantable cardioverter-defibrillator; defibrillator shock; cancelled defibrillator therapy; remote monitoring

向作者/读者索取更多资源

Background: Cancelled shock therapy (CxTx) may presage shock delivery and shorten battery longevity of implantable cardioverter-defibrillators (ICDs). However, it is silent and has received little attention. Remote home monitoring (HM) with continuous surveillance and automatic daily data archiving allows investigation of CxTx. Methods and Results: We retrospectively analyzed a database of 4,960 recipients of HM ICD, from 2002 to 2007. Over a mean follow-up of 445 +/- 253 days, CxTx occurred in 1,392 (28%) patients, mostly as single episodes (n = 1,120). However, 142 patients (10% of patients with CxTx, 2.9% of the whole ICD population) had >10 CxTx. CxTx was followed by shock delivery in 432 patients, in 239 (55%) of whom CxTx occurred >10 days prior to shock delivery. In 113 patients (26%), CxTx occurred in the 72-hour period preceding the shock. A single CxTx preceded the shock in 74 of these 113 patients, and only eight patients had >3 CxTx. Conclusion: CxTx was highly prevalent in unselected ICD recipients, though the intraindividual density of episodes was low. In 2.9% of the overall population, the number of CxTx was sufficient to shorten the battery longevity. A high number of CxTx usually did not predict the delivery of shocks. (PACE 2009; 32:S42-S46)

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据