Corneal Deformation Measurement Using Scheimpflug Noncontact Tonometry

期刊

OPTOMETRY AND VISION SCIENCE
卷 90, 期 1, 页码 E1-E8

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/OPX.0b013e318279eb87

关键词

noncontact tonometry; repeatability; reproducibility; corneal deformation amplitude; corneal biomechanics

向作者/读者索取更多资源

Purpose. To determine the intraexaminer repeatability and intersession reproducibility of corneal deformation measurement using Scheimpflug noncontact tonometry (Corvis ST) on normal subjects. Methods. Thirty-seven adults aged 20 to 48 years were invited to have their corneal deformation and curvature measurements taken using Corvis ST and Pentacam, respectively. Three consecutive measurements were taken for each instrument between 9:00 and 11:00 AM for intraexaminer repeatability analysis. Participants returned between 3:00 and 5:00 PM the same day for intersession reproducibility analysis. Results. The most repeatable corneal parameter measured by Corvis ST was central corneal thickness ([CCT] ICC, 0.96; precision, 10.85 mu m; repeatability, 15.34 mu m; CV, 1.01%), followed by deformation amplitude ([DA] ICC, 0.80; precision, 0.08 mm; repeatability, 0.13 mm; CV, 4.33%), first applanation time ([1st A-time] ICC, 0.77; precision, 0.22 milliseconds; repeatability, 0.31 milliseconds; CV, 1.42), and intraocular pressure ([IOP] ICC, 0.75; precision, 1.39 mm Hg; repeatability, 1.97 mm Hg; CV, 4.98). Other parameters showed poor repeatability. The DA and 1st A-time showed good intersession reproducibility. The 95% limits of agreement were +0.13 to (-)0.13 mm for DA and +0.27 to (-)0.33 milliseconds for 1st A-time. The DA was negatively correlated with central corneal thickness (r = (-)0.53, p < 0.001) but not with corneal curvatures (flattest curvature, r = 0.13, p = 0.46; steepest curvature, r = 0.05, p = 0.75). Conclusions. Corneal deformation parameters DA and 1st A-time were repeatable and reproducible. A thinner cornea was associated with a higher corneal deformation. Measurement of DA serves as an indicator of corneal biomechanical properties. (Optom Vis Sci 2013;90:e1-e8)

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据