A Novel On-Eye Wettability Analyzer for Soft Contact Lenses

期刊

OPTOMETRY AND VISION SCIENCE
卷 88, 期 10, 页码 E1188-E1195

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/OPX.0b013e318227e30d

关键词

contact lens; in vivo; wettability; spreading; hydrogel

资金

  1. Jordan University of Science and Technology

向作者/读者索取更多资源

Purpose. To develop novel methodology to assess two in vivo wettability parameters, contact angle and index of liquid spreading (ILS), for hydrogel contact lenses and to provide a comprehensive investigation of the measurement errors (repeatability) associated with these two parameters. Methods. A Novel On-eye Wettability Analyzer (NOWA) was constructed which delivered a drop of 0.4% sodium hyaluronate mixed with sodium fluorescein directly on to a lens surface in vivo while a two-camera digital system recorded both the resultant contact angle and liquid spreading. Ten subjects each wore a different etafilcon A lens (-0.50 DS) in the right eye for 30 min before measurement on three separate occasions. For contact angles, the coefficient of repeatability (COR) associated with three measurement conditions (image analysis COR, intrasubject COR, and intersubject COR) was determined. For ILS, the COR associated with two measurement conditions (intrasubject COR and inter-subject COR) was also determined. Results. The in vivo contact angle for the etafilcon A lens was 19.8 degrees +/- 2.9 degrees (mean +/- standard deviation). COR values for contact angle were 3.2 degrees, 5.2 degrees, and 8.0 degrees for image analysis, intrasubject measures, and intersubject measures, respectively. The mean +/- standard deviation ILS was 2.0 +/- 0.3 mm, and COR values were 0.4 and 0.8 mm for intrasubject and intersubject measures, respectively. Conclusions. This research demonstrates that it is possible to obtain hydrogel lens contact angles in vivo and to provide a new wettability parameter (ILS) with the NOWA instrumentation. The COR values obtained were considered to be acceptable. Intersubject COR values were greater than intrasubject COR values for both contact angle and ILS; this is likely to be due to greater variability of tear film characteristics between subjects than within an individual. The use of this novel methodology and its associated clinical parameters across a wider range of lens types in various clinical situations warrants further investigation. (Optom Vis Sci 2011; 88: E1188-E1195)

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据