期刊
OPTICS EXPRESS
卷 22, 期 12, 页码 14246-14252出版社
OPTICAL SOC AMER
DOI: 10.1364/OE.22.014246
关键词
-
类别
Grating-based x-ray differential phase contrast imaging (DPCI) often uses a phase stepping procedure to acquire data that enables the extraction of phase information. This method prolongs the time needed for data acquisition by several times compared with conventional x-ray absorption image acquisitions. A novel analyzer grating design was developed in this work to eliminate the additional data acquisition time needed to perform phase stepping in DPCI. The new analyzer grating was fabricated such that the linear grating structures are shifted from one detector row to the next; the amount of the lateral shift was equal to a fraction of the x-ray diffraction fringe pattern. The x-ray data from several neighboring detector rows were then combined to extract differential phase information. Initial experimental results have demonstrated that the new analyzer grating enables accurate DPCI signal acquisition from a single x-ray exposure like conventional x-ray absorption imaging. (C) 2014 Optical Society of America
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐