4.6 Article

Towards a portable Raman spectrometer using a concave grating and a time-gated CMOS SPAD

期刊

OPTICS EXPRESS
卷 22, 期 15, 页码 18736-18747

出版社

OPTICAL SOC AMER
DOI: 10.1364/OE.22.018736

关键词

-

类别

向作者/读者索取更多资源

A low-cost, compact Raman spectrometer suitable for the on-line water monitoring applications is explored. A custom-designed concave grating for wavelength selection was fabricated and tested. The detection of the Raman signal is accomplished with a time-gated single photon avalanche diode (TG-SPAD). A fixed gate window of 3.5ns is designed and applied to the TG-SPAD. The temporal resolution of the SPAD was similar to 60ps when tested with a 7ps, 532nm solid-state laser. To test the efficiency of the gating in fluorescence signal suppression, different detection windows (3ns-0.25ns) within the 3.5ns gate window are used to measure the Raman spectra of Rhodamine B. Strong Raman peaks are resolved with this low-cost system. (C)2014 Optical Society of America