4.6 Article

Integrating fault tolerance algorithm and circularly polarized ellipsometer for point-of-care applications

期刊

OPTICS EXPRESS
卷 19, 期 6, 页码 5431-5441

出版社

OPTICAL SOC AMER
DOI: 10.1364/OE.19.005431

关键词

-

类别

资金

  1. Taiwan National Science Council (NSC) [NSC 98-2627-E-002-003]

向作者/读者索取更多资源

A circularly polarized ellipsometer was developed to enable real-time measurements of the optical properties of materials. Using a four photo-detector quadrature configuration, a phase modulated ellipsometer was substantially miniaturized which has the ability to achieve a high precision detection limit. With a proven angular resolution of 0.0001 deg achieved by controlling the relative positions of a triangular prism, a paraboloidal and a spherical mirror pair, this new ellipsometer possesses a higher resolution than traditional complex mechanically controlled configurations. Moreover, the addition of an algorithm, FTA (fault tolerance algorithm) was adopted to compensate for the imperfections of the opto-mechanical system which can decrease system measurement reliability. This newly developed system requires only one millisecond or less to complete the measurement task without having to adopt any other modulation approach. The resolution achieved can be as high as 4x10(-7) RIU (refractive index unit) which is highly competitive when compared with other commercially available instruments. Our experimental results agreed well with the simulation data which confirms that our quadrature-based circularly polarized ellipsometer with FTA is an effective tool for precise detection of the optical properties of thin films. It also has the potential to be used to monitor the refractive index change of molecules in liquids. (C)2011 Optical Society of America

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据