4.6 Article Proceedings Paper

Positron emission mammography using Pr:LuAG scintillator - Fusion of optical material study and systems engineering

期刊

OPTICAL MATERIALS
卷 32, 期 10, 页码 1294-1297

出版社

ELSEVIER
DOI: 10.1016/j.optmat.2010.05.002

关键词

Pr:LuAG; PET; PEM; Scintillator; Gamma camera

向作者/读者索取更多资源

The breast cancer is one of the most frequent cause of death among female cancer patients. However, when discovered at the early stage, the probability of recovery is very high. Therefore, the development of positron emission mammography (PEM) to detect the breast cancer at the early stage with high efficiency is demanding. As the diameter of the scanner part is small, a scintillator with faster response is required. As the recently developed Pr:LuAG has almost twice as short scintillation decay time than that of Ce:LSO, the Pr:LuAG scintillator was employed in our recently developed PEM system. One camera unit consisted of 20 x 64 scintillator pixels optically coupled with three H8500-03 multi anode photomultipliers. The Pr:LuAG pixel size is 2.1 x 2.1 x 15 mm(3) and the BaSO4 was used as a reflector. Four planar cameras are placed at both sides. Therefore, eight cameras were installed at both sides of the instrument. The spatial resolution was evaluated to be 1.1 mm using the Na-22 point source. Fluorodeoxyglucose with F-18 hotspot image was also detected using the breast phantom. (C) 2010 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据