A disposable electrochemical sensor based on protein G for High-Density Lipoprotein (HDL) detection

期刊

TALANTA
卷 144, 期 -, 页码 466-473

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2015.06.009

关键词

Protein G; HDL; AFM; Contact angle; Impedance spectroscopy

向作者/读者索取更多资源

In this work, two biosensors were developed for the detection of High-Density Lipoproteins (HDL) particles, which are biomarkers inversely correlated with cardiovascular risk and which represent therapeutic targets for atherosclerosis. The electrochemical properties of the grafted antibody on interdigitated gold electrode were achieved by Impedance Spectroscopy (IS). The used deposition method was based on oriented antibody Anti-ApoA1 with an intermediate thin layer of protein G. The developed biosensor was able to detect both native plasma HDL and reconstituted HDL (rHDL) particles respectively with the detection limit of 50 ng/mL and 1 ng/mL, respectively. Dynamic contact angle and atomic force microscopy were used. The developed biosensors are able to differentiate the HDL particles according to their differences in size and interactions with the immobilized antibody. (C) 2015 Elsevier B.V. All rights reserved.