4.7 Article

Determination of the maleic acid in rat urine and serum samples by isotope dilution-liquid chromatography-tandem mass spectrometry with on-line solid phase extraction

期刊

TALANTA
卷 136, 期 -, 页码 9-14

出版社

ELSEVIER
DOI: 10.1016/j.talanta.2014.11.021

关键词

Maleic acid; Serum; Urine; Isotope dilution method; On-line solid-phase extraction; Liquid chromatography-tandem mass spectrometry

资金

  1. National Science Council of Taiwan
  2. National Taiwan University

向作者/读者索取更多资源

A rapid and simple on-line solid-phase extraction coupled with isotope dilution-liquid chromatography-tandem mass spectrometry (SPE-ID-LC-MS/MS) method was developed to quantitate maleic acid in serum and urine of SpragueDawley (SD) rats. The aforementioned biological samples were spiked with C-13(2)-maleic acid, vigorously vortexed, added with acetonitrile to precipitate proteins, and then injected into the on-line SPE-LC-MS/MS system for quantification. Upon validation, this method demonstrated excellent feasibility and sensitivity: calibration curves for maleic acid in serum and urine display excellent linearity with the coefficient of determination (R-2) greater than 0.999; the limits of detection and quantitation (LOD and LOQ) for maleic acid were determined at 0.2 and 0.5 mu g L-1, respectively. Additionally, intra-day accuracy for maleic acid in serum and urine samples ranged from 94.0% to 100.2% and 101.3% to 104.4%, respectively. Furthermore, inter-day accuracy ranged from 93.6% to 101.0% and from 102.3% to 111.4% in serum and urine samples, respectively. Intra-day precision %RSD of maleic acid in serum and urine samples was 13.8% or less, whereas the inter-day precision was 6.1% or less. The matrix effects were not found to be statistically significant (p=0.9145 and p=0.5378, correspondingly) based on the calculations of recovery functions. The collected serum and urine samples were analyzed using SPE-ID-LC-MS/MS. Our results reveal trace levels of maleic acid in the control rats, demonstrating that this method is capable of analyzing background levels of contaminants in biofluids with excellent sensitivity and specificity at part-per-billion levels concentrations in complex matrices. (C) 2015 Published by Elsevier B.V.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据