期刊
NUCLEAR TECHNOLOGY & RADIATION PROTECTION
卷 33, 期 4, 页码 363-368出版社
VINCA INST NUCLEAR SCI
DOI: 10.2298/NTRP180316004V
关键词
ethanol-chlorobenzene; oscillotitrator; dosimetry; radiation
资金
- Ministry of Education, Science and Technological Development of the Republic of Serbia [45020]
Ethanol-chlorobenzene/oscillotitrator dosimetry system is widely used in controlling the irradiation process in gamma facilities. The ethanol-chlorobenzene dosimetry system provides a reliable means of measuring absorbed dose. It is based on a process of radiolytic formation of hydrochloric acid in aqueous ethanolic solutions of chlorobenzene by ionizing radiation. The irradiation temperature dependence of dosimeter response is a complex function of dose and temperature for each concentration of chlorobenzene. At different temperature the mobility of conducting species from hydrochloric acid is changed leading to different oscillotitrator deflections during high-frequency conductometric readout. In this paper, we examined the influence of temperature on the calculation of the radiation dose. We showed that the temperature significantly influenced the measurement results, and that the calibration curve has to be formed at the irradiation temperature in order to obtain precise values of the absorbed dose.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据