Spectral X-ray imaging; Mammography; Silicon strip detector; Photon counting; Energy resolution; Cascaded detector model
类别
资金
- European Union
A photon-counting silicon strip detector with two energy thresholds was investigated for spectral X-ray imaging in a mammography system. Preliminary studies already indicate clinical benefit of the detector, and the purpose of the present study is optimization with respect to energy resolution. Factors relevant for the energy response were measured, simulated, or gathered from previous studies, and used as input parameters to a cascaded detector model. Threshold scans over several X-ray spectra were used to calibrate threshold levels to energy, and to validate the model. The energy resolution of the detector assembly was assessed to range over Delta E/E=0.12-0.26 in the mammography region. Electronic noise dominated the peak broadening, followed by charge sharing between adjacent detector strips, and a channel-to-channel threshold spread. The energy resolution may be improved substantially if these effects are reduced to a minimum. Anti-coincidence logic mitigated double counting from charge sharing, but erased the energy resolution of all detected events, and optimization of the logic is desirable. Pile-up was found to be of minor importance at typical mammography rates. (C) 2009 Elsevier B.V. All rights reserved.