NMR IN BIOMEDICINE
卷 23, 期 9, 页码 1103-1108出版社
WILEY
DOI: 10.1002/nbm.1526
关键词
motion correction; real time; MRI; ultrasound echo
资金
- Ligue Nationale Contre le Cancer
- Conseil Regional d Aquitaine Diagnostic Molecular Imaging
- EC [LSHB-CT 2005 512146]
- Agence National de la Recherche
- Philips Healthcare
The objective of this study was to evaluate the feasibility of integrating real time ultrasound echo guidance in MR-guided high-intensity focused ultrasound (HIFU) heating of mobile targets in order to reduce latency between displacement analysis and HIFU treatment Experiments on a moving phantom were carried out with MRI-guided HIFU during continuous one-dimensional ultrasound echo detection using separate HIFU and ultrasound imaging transducers Excellent correspondence was found between MR- and ultrasound-detected displacements Real-time ultrasound echo-based target tracking during MR-guided HIFU heating is shown with the dimensions of the heated area similar to those obtained for a static target This work demonstrates that the combination of the two modalities opens up perspectives for motion correction in MRI-guided HIFU with negligible latency Copyright (C) 2010 John Wiley & Sons, Ltd