期刊
NEUROSURGERY
卷 68, 期 4, 页码 1131-1143出版社
OXFORD UNIV PRESS INC
DOI: 10.1227/NEU.0b013e318209cc45
关键词
Application accuracy; Image-guided neurosurgery; Neuronavigation
资金
- Science and Technology Commission of Shanghai Municipality [08dz1900705, 09410702800]
- National High Technology Research and Development Program [2009AA02Z415]
- Shanghai Leading Academic Discipline Project [B112]
There are many different types of errors in neuronavigation, and the reasons and results of these errors are complex. For a neurosurgeon using the neuronavigation system, it is important to have a clear understanding of when an error may occur, what the magnitude of it is, and how to avoid it or reduce its influence on the final application accuracy. In this article, we classify all the errors into 2 groups according to the working principle of neuronavigation systems. The first group contains the errors caused by the differences between the anatomic structures in the images and that of the real patient, and the second group contains the errors occurring in transforming the position of surgical tools from the patient space to the image space. Each group is further divided into 2 subgroups. We discuss 16 types of errors and classify each of them into one of the subgroups. The classification and analysis of these errors should help neurosurgeons understand the power and limits of neuronavigation systems and use them more properly.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐