THE NEW LICOX COMBINED BRAIN TISSUE OXYGEN AND BRAIN TEMPERATURE MONITOR: ASSESSMENT OF IN VITRO ACCURACY AND CLINICAL EXPERIENCE IN SEVERE TRAUMATIC BRAIN INJURY

期刊

NEUROSURGERY
卷 63, 期 6, 页码 1159-1164

出版社

OXFORD UNIV PRESS INC
DOI: 10.1227/01.NEU.0000333265.19131.7C

关键词

Brain temperature; Brain tissue oxygen tension; Clinical evaluation; In vitro accuracy; Technology assessment

向作者/读者索取更多资源

OBJECTIVE: Monitoring of brain tissue oxygen tension is increasingly being used to monitor patients after severe traumatic brain injury and to guide therapies aimed at maintaining brain tissue oxygen tension above threshold levels. The new Licox PMO combined oxygen and temperature catheter (Integra LifeSciences, Plainsboro, NJ) combines measurements of oxygen tension and temperature in a single probe inserted through a bolt mechanism. In this study, we sought to evaluate the accuracy of the new Licox PMO probe under controlled laboratory conditions and to assess the accuracy of oxygen tension and temperature measurements and the new automated card calibration system. We also describe our clinical experience with the Licox PMO probe. METHODS: Oxygen tension was measured in a 2-chambered apparatus at different oxygen tensions and temperatures. The new card calibration system was compared with a manually calibrated system. Rates of hematoma, infection, and dislodgement in Our clinical experience were recorded. RESULTS: The new Licox PMO probe accurately measures oxygen tension over a wide range of oxygen concentrations and physiological temperatures, but it does have a small tendency to underestimate oxygen tension (mean error, -3.8 +/- 3.5%) that is more pronounced between the temperatures of 33 and 39 degrees C. The thermistor of the PMO probe also has a tendency to underestimate temperature when compared with a resistance thermometer (mean error, -0.67 +/- 0.22 degrees C). The card calibration system was also found to introduce some variability in measurements of oxygen tension when compared with a manually calibrated system. Clinical experience with the new probe indicates good placement within the white matter using the improved bolt system and low rates of hematoma (2.9%), infection (0%), and dislodgement (5.9%). CONCLUSION: The new Licox PMO probe is accurate but has a small, consistent tendency to under-read oxygen tension that is more pronounced at higher temperatures. The probe tends to under-read temperature by 0.5 to 0.8 degrees C across temperatures, suggesting that caution should be used when brain temperature is measured with the Licox PMO probe and used to guide tempera tu re-directed treatment strategies. The Licox PMO probe improves upon previous models in allowing consistent and accurate placement in the white matter and obviating the need for placement of 2 separate probes to measure oxygen tension and temperature.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据