期刊
SENSORS AND ACTUATORS B-CHEMICAL
卷 210, 期 -, 页码 254-258出版社
ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2014.12.108
关键词
Amperometry; Benzylpenicillin; Carbon paste biosensor; Differential pulse voltammetry; Electrochemistry; Penicillinase
资金
- FAPESP [2011/17552-0]
- Fundacao para a Ciencia e a Tecnologia (FCT) [SFRH/BPD/76544/2011]
- Fundacao de Amparo a Pesquisa do Estado de Sao Paulo (FAPESP) [11/17552-0] Funding Source: FAPESP
Antibiotics are widely used in dairy cattle production, however this can pose a public health threat, particularly when antibiotic residues remain in the food products like milk. Therefore, new analytical solutions aiming the determination of antibiotics, like the still commonly used penicillin, are welcomed. This work describes a novel electrochemical biosensor, with amperometric and differential pulse voltammetric detection, based on a modified carbon paste with beta-lactamase for the determination of penicillin G, a beta-lactam antibiotic. It is a 'second-generation' biosensor since cobalt phthalocyanine was used as an electron mediator. The proposed biosensor, by differential pulse voltammetry, presented a limit of detection of 0.079 mu mol/L. Using this biosensor, a value of 43 +/- 3 mu mol/L was determined for the Michaelis-Menten enzymatic constant for penicillinase extracted from Bacillus cereus. (C) 2015 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐