4.7 Article

Ambiguities in applying traditional Limulus Amebocyte Lysate tests to quantify endotoxin in nanoparticle formulations

期刊

NANOMEDICINE
卷 5, 期 4, 页码 555-562

出版社

FUTURE MEDICINE LTD
DOI: 10.2217/NNM.10.29

关键词

endotoxin; in vitro assay; interference; Limulus Amebocyte Lysate; lipopolysaccharide; nanoparticles; rabbit pyrogen test

资金

  1. National Cancer Institute, NIH [N01-CO-12400, HHSN261200800001E]
  2. NIH/NCI R01 [CA 123495]

向作者/读者索取更多资源

Nanotechnology is finding increasing application in biology and medicine. As with other pharmaceutical formulations and medical devices intended for use in animals and human patients, contamination of nanoparticles with bacterial endotoxins should be thoroughly investigated before preclinical in vitro and in vivo characterization. Traditional methods to study endotoxin contamination include the in vitro quantitative Limulus amebocyte lysate test and the in vivo qualitative rabbit pyrogen test. Both of these tests have a long history of use for traditional pharmaceuticals and medical devices and are routinely used in drug development. Here we report that nanoparticles often interfere with these traditional endotoxin detection tests and suggest approaches to detect and overcome such interferences.