Evaluation of the Litron In Vitro MicroFlow® Kit for the flow cytometric enumeration of micronuclei (MN) in mammalian cells

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.mrgentox.2008.05.003

关键词

genotoxicity testing; micronuclei; flow cytometry; cytotoxicity

向作者/读者索取更多资源

We have evaluated the performance of the prototype In Vitro MicroFlow((R)) Kit (Litron Laboratories), which offers a flow cytometric method for scoring micronuclei (MN). This method uses sequential staining to differentiate MN from chromatin fragments derived from apoptotic or necrotic cells. Data were generated using the genotoxins methylmethane sulphonate (MMS), dimethylbenzanthracene (DMBA) and vinblastine, and the non-genotoxins dexamethasone and staurosporine, which are known to induce apoptosis in vitro. The results obtained with these agents were compared with conventional microscopy. For short-duration exposures (3-4 h) both manual and flow methodologies demonstrated good concordance, with concentration-related increases in the percentage of MN for MMS, DMBA and vinblastine. Statistically significant increases were observed at >= 20 and 40 mu g/mL, for manual and flow analysis, respectively, for MMS; at 0.5 and 0.75 mu g/mL for DMBA: and at 0.035 and 0.04 mu g/mL, respectively, for vinblastine. Dexamethasone showed clear negative responses by manual and flow cytometric analysis, with comparable results for both methodologies (all < 1.7-fold compared with concurrent vehicle controls). Data for staurosporine, however, were less consistent showing significantly higher flow cytometric MN frequencies compared with those seen after manual analysis. Continuous (24h) treatments were also conducted with MMS, vinblastine, dexamethasone and staurosporine. There was good concordance between the methodologies for MMS, staurosporine and vinblastine. However, dexamethasone generated discordant results, i.e. microscopic analysis was clearly negative at all doses tested, whereas flow cytometry produced significant increases in MN frequency (up to 8.1-fold at 100 mu g/mL compared with the concurrent vehicle control). The inconsistencies observed between flow cytometry and standard microscopy, and the differences in assay sensitivity, particularly for apoptosis-inducing compounds, suggest that the prototype In Vitro MicroFlow((R)) Kit requires further refinement. Studies to investigate new parameters to address these issues are now under way and will be reported separately. (c) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据