期刊
SENSORS
卷 15, 期 6, 页码 12668-12681出版社
MDPI
DOI: 10.3390/s150612668
关键词
whole cell biosensor; hazardous heavy metal; optical fiber biosensor; GFP E. coli
资金
- Universiti Kebangsaan Malaysia [DPP-2015, DIP-2014-016]
A fluorescence-based fiber optic toxicity biosensor based on genetically modified Escherichia coli (E. coli) with green fluorescent protein (GFP) was developed for the evaluation of the toxicity of several hazardous heavy metal ions. The toxic metals include Cu(II), Cd(II), Pb(II), Zn(II), Cr(VI), Co(II), Ni(II), Ag(I) and Fe(III). The optimum fluorescence excitation and emission wavelengths of the optical biosensor were 400 +/- 2 nm and 485 +/- 2 nm, respectively. Based on the toxicity observed under optimal conditions, the detection limits of Cu(II), Cd(II), Pb(II), Zn(II), Cr(VI), Co(II), Ni(II), Ag(I) and Fe(III) that can be detected using the toxicity biosensor were at 0.04, 0.32, 0.46, 2.80, 100, 250, 400, 720 and 2600 mu g/L, respectively. The repeatability and reproducibility of the proposed biosensor were 3.5%-4.8% RSD (relative standard deviation) and 3.6%-5.1% RSD (n = 8), respectively. The biosensor response was stable for at least five weeks, and demonstrated higher sensitivity towards metal toxicity evaluation when compared to a conventional Microtox assay.