4.4 Article

ELECTROMAGNETIC VENTILATION: FIRST EVALUTION OF A NEW METHOD FOR ARTIFICAL VENTILATION IN HUMANS

期刊

MUSCLE & NERVE
卷 42, 期 3, 页码 305-310

出版社

JOHN WILEY & SONS INC
DOI: 10.1002/mus.21698

关键词

electromagnetic stimulation; electromagnetic ventilation; mechanical ventilation; respiratory muscles; respiratory muscle atrophy

向作者/读者索取更多资源

Current methods of artificial ventilation cannot prevent diverse problems associated with mechanical ventilation. In contrast to all current forms of mechanical ventilation, electromagnetic stimulation can activate respiratory muscles directly. However, it is not known if and to what extent electromagnetic stimulation can ventilate humans. In 10 volunteers we stimulated the lateral neck using magnetic stimulators. Over 63 s we stimulated nine times with a frequency of 25 Hz for 1.1 s using 600 V, 900 V, and 1,200 V. The minimum stimulation time for each volunteer was 9 min. A Capnomac monitor measured minute ventilation. Electromagnetic stimulation was well tolerated and safe. Bilateral stimulation with 600 V achieved considerable minute ventilation (median +/- SD, 7.2 +/- 3.4 L/min) that increased at higher voltage levels (P < 0.0001). 900 V achieved sufficient minute ventilation in all volunteers (11.5 +/- 5.0 L/min; 1200 V, 14.0 +/- 4.9 L/min). This first evaluation of electromagnetic ventilation demonstrates that it can be used to ventilate humans sufficiently. This method may be developed to a new mode of ventilation. Muscle Nerve 42: 305-310, 2010