4.5 Article

Low-Voltage Electron-Probe Microanalysis of Fe-Si Compounds Using Soft X-Rays

期刊

MICROSCOPY AND MICROANALYSIS
卷 19, 期 6, 页码 1698-1708

出版社

CAMBRIDGE UNIV PRESS
DOI: 10.1017/S1431927613012695

关键词

soft X-ray; low voltage; nanoanalysis; microprobe; EPMA

向作者/读者索取更多资源

Conventional electron-probe microanalysis has an X-ray analytical spatial resolution on the order of 1-4 mu m width/depth. Many of the naturally occurring Fe-Si compounds analyzed in this study are smaller than 1 mu m in size, requiring the use of lower accelerating potentials and nonstandard X-ray lines for analysis. Problems with the use of low-energy X-ray lines (soft X-rays) of iron for quantitative analyses are discussed and a review is given of the alternative X-ray lines that may be used for iron at or below 5 keV (i.e., accelerating voltage that allows analysis of areas of interest <1 mu m). Problems include increased sensitivity to surface effects for soft X-rays, peak shifts (induced by chemical bonding, differential self-absorption, and/or buildup of carbon contamination), uncertainties in the mass attenuation coefficient for X-ray lines near absorption edges, and issues with spectral resolution and count rates from the available Bragg diffractors. In addition to the results from the traditionally used Fe L alpha line, alternative approaches, utilizing Fe L beta, and Fe Ll-eta lines, are discussed.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据