期刊
MICRON
卷 43, 期 11, 页码 1147-1155出版社
PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.micron.2012.01.007
关键词
In situ TEM; Environmental (scanning) transmission electron microscope (E(S)TEM); Catalysis; Nanowires; Nanotubes
类别
Recently, the applications of transmission electron microscopy (TEM) related techniques have extended from ex situ nanoscale characterization of structure and chemistry of products to dynamic measurements of nanostructures during reaction processes. Commercially available modified TEM specimen holders and TEM columns are being routinely employed to follow the structural and chemical changes at elevated temperatures and even under controlled atmosphere. Experiments performed under these rigorous conditions require careful considerations to avoid undesirable effects from the gas impurities or contaminations from TEM grids and/or holders. The reactivity of sample, grid, holder, TEM components, and gaseous environments must be evaluated for each reaction process. This tutorial is aimed to outline some of the important factors that should be considered for experimental set up used for in situ observations to ensure the results are comparable to the ones obtained during ex situ experiments under identical conditions. Published by Elsevier Ltd.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐