4.7 Article

Gold nanoparticle-coated multiwall carbon nanotube-modified electrode for electrochemical determination of methyl parathion

期刊

MICROCHIMICA ACTA
卷 175, 期 3-4, 页码 309-314

出版社

SPRINGER WIEN
DOI: 10.1007/s00604-011-0681-5

关键词

Gold nanoparticles; Carbon nanotubes; Methyl parathion; Electrochemical sensor

资金

  1. Ministry of Science and Technology of China [2008AA06Z311]

向作者/读者索取更多资源

We report on an electrochemical sensor for the determination of methyl parathion. It is based on an electrode modified with multi-walled carbon nanotubes that were covered with gold nanoparticles (Au-NPs). The vertically aligned array of MWCNTs on a tantalum substrate was coated with Au-NPs by overhead magnetron sputtering deposition. Scanning and transmission electron microscopy and XRD were used to characterize the Au-NP-MWCNTs composite. Cyclic voltammetry and differential pulse voltammetry were employed to evaluate the suitability of the new electrode for the determination of methyl parathion. Under the optimal conditions, the current response of the electrode to methyl parathion is linear in the range from 0.50 to 16.0 mg mL(-1), with a detection limit of 50 mu g mL(-1) (signal/noise = 3), and the sensitivity is 4.5 times better than that of the plain MWCNTs electrode. We conclude that this method represents a simple, rapid, effective and sensitive approach for the detection of methyl parathion.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据