4.7 Article

Electrodeposition of silver nanoparticles on a zinc oxide film: improvement of amperometric sensing sensitivity and stability for hydrogen peroxide determination

期刊

MICROCHIMICA ACTA
卷 169, 期 3-4, 页码 361-365

出版社

SPRINGER WIEN
DOI: 10.1007/s00604-010-0356-7

关键词

Modified electrode; Non-enzymatic sensors; Silver nanoparticles; Zinc oxide; Hydrogen peroxide

资金

  1. National Natural Science Foundation of China [20875076]

向作者/读者索取更多资源

A strategy to fabricate a hydrogen peroxide (HP) sensor is developed by electrodepositing silver nanoparticles (Ag NPs) on a modified glassy carbon electrode (GCE) with a zinc oxide (ZnO) film. The Ag NPs/ZnO/GCE has been characterized by scanning electron microscopy, cyclic voltammetry, and chronoamperometry. It has been found that the Ag NPs synthesized in the presence of ZnO film provide an electrode with enhanced sensitivity and excellent stability. The sensitivity to HP is enhanced 3-fold by using Ag NPs/ZnO/GCE compared to Ag NPs/GCE. The HP sensor exhibits good linear behavior in the concentration range 2 A mu M to 5.5 mM for the quantitative analysis of HP with a detection limit of 0.42 A mu M (S/N = 3).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据