Flame atomic absorption spectrometric determination of trace amounts of Pb(II) and Cr(III) in biological, food and environmental samples after preconcentration by modified nano-alumina

期刊

MICROCHIMICA ACTA
卷 172, 期 1-2, 页码 125-136

出版社

SPRINGER WIEN
DOI: 10.1007/s00604-010-0478-y

关键词

Preconcentration; Modified gamma-alumina nanoparticles; Lead (II); Chromium (III); Atomic absorption spectrometric determination

资金

  1. Bu-Ali Sina University Research Council
  2. Center of Excellence in Development of Chemical Methods (CEDCM)

向作者/读者索取更多资源

A new solid-phase extraction sorbent was used for the preconcentration of Pb(II) and Cr(III) ions prior to their determination by flame atomic absorption spectrometry. It was prepared by immobilization of 2,4-dinitrophenylhydrazine on nano-alumina coated with sodium dodecyl sulfate. The sorbent was characterized by scanning electron microscopy, N-2 adsorption and Fourier transform infrared spectrometry, and used for preconcentration and separation of Pb(II) and Cr(III) from aqueous solutions. The ions on the sorbent were eluted with a mixture of nitric acid and methanol. The effects of sample pH, flow rates of samples and eluent, type of eluent, breakthrough volume and potentially interfering ions were studied. Linearity is maintained between 1.2 and 350 mu g L-1 of Pb(II), and between 2.4 and 520 mu g L-1 of Cr(III) for an 800-mL sample. The detection limit (3 s, N = 10) for Pb(II) and Cr(III) ions is 0.43 and 0.55 mu g L-1, respectively, and the maximum preconcentration factor is 267. The method was successfully applied to the evaluation of these trace and toxic metals in various water, food, industrial effluent and urine samples.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据