In this work, inductively coupled plasma emission spectroscopy was utilized to directly determine the chemical composition of magnetic fluids constituted of size-sorted ferrite nanoparticles in aqueous solution. Nickel and cobalt nanoferrites were chemically synthesized following a bottom-up route and dispersed under various pH conditions. Size and structural characteristics of nanograins were investigated by X-ray diffraction using a synchrotron source. Chemical analysis was then carried out by directly introducing diluted magnetic fluid samples (slurries) into the spectrometer. To achieve reliable measurements, sample conditions and apparatus parameters were carefully investigated. Slurry stability must be optimized in order to obtain reproducible and accurate analysis. The instrument must also be calibrated to minimize the difference between the signal produced by slurries and that of aqueous ordinary solutions. Furthermore, slurry sample introduction offers many advantages over conventional sample digestion, including reduced sample pretreatment time, less possibility of contamination and the use of direct calibration with aqueous solutions. (C) 2010 Elsevier B.V. All rights reserved.