期刊
METROLOGIA
卷 48, 期 5, 页码 283-289出版社
IOP PUBLISHING LTD
DOI: 10.1088/0026-1394/48/5/007
关键词
-
资金
- NSF
- Penn State
- UK National Measurement Office
We evaluate the distributed cavity phase (DCP) and microwave lensing frequency shifts, which were the two largest sources of uncertainty for the NPL-CsF2 caesium fountain clock. We report measurements that confirm a detailed theoretical model of the microwave cavity fields and the frequency shifts of the clock that they produce. The model and measurements significantly reduce the DCP uncertainty to 1.1 x 10(-16). We derive the microwave lensing frequency shift for a cylindrical cavity with circular apertures. An analytic result with reasonable approximations is given, in addition to a full calculation that indicates a shift of 6.2 x 10(-17). The measurements and theoretical models we report, along with improved evaluations of collisional and microwave leakage induced frequency shifts, reduce the frequency uncertainty of the NPL-CsF2 standard to 2.3 x 10(-16), nearly a factor of two lower than its most recent complete evaluation.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐