4.3 Article

Vibrations in the human middle ear

期刊

MEDICAL SCIENCE MONITOR
卷 17, 期 12, 页码 BR372-BR376

出版社

INT SCIENTIFIC INFORMATION, INC
DOI: 10.12659/MSM.882123

关键词

middle ear vibrations; laser doppler vibrometer; nonlinear dynamics

资金

  1. Polish Ministry of Science and Higher Education [N403 065 32/3451, N N518 425936]

向作者/读者索取更多资源

Background: Middle ear surgery techniques can improve hearing destroyed by disease, but results of treatment are difficult to predict. Therefore, researchers use a Laser Doppler Vibrometer to measure vibrations of human middle ear ossicles. Material/Methods: Measurements of ossicular chain vibrations are performed on fresh human temporal bone specimens using Laser Doppler Vibrometer. Vibrations of stapes are recorded in 3 cases: 1) for intact ossicular chain, 2) when incus long process is removed, and 3) after long process reconstruction with bone cement. A typical analysis of transfer function is completed by other methods applied in dynamics. Results: Measurements and analysis of stapes vibrations in case of intact and damaged ossicular chain show regular and irregular behavior which can be recognize with the help of phase portraits, recurrence plots, correlation dimension, and Hurst and Lyapunov exponents. The long process reconstruction with bone cement gives good results in improving hearing. Conclusions: Recurrence plots, and Lyapunov and Hurst exponents used in the study complete information obtained from transfer function and can be employed to enrich the classical approach to ossicular chain vibrations.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据