4.6 Article

The effect of the flattening filter on photoneutron production at 10MV in the Varian TrueBeam linear accelerator

期刊

MEDICAL PHYSICS
卷 45, 期 10, 页码 4711-4719

出版社

WILEY
DOI: 10.1002/mp.13148

关键词

10MV; flattening filter-free; MLEM; neutron; TrueBeam

资金

  1. Natural Sciences and Engineering Research Council of Canada (NSERC)
  2. CREATE Medical Physics Research Training Network grant of the Natural Sciences and Engineering Research Council [432290]
  3. Canadian Nuclear Safety Commission

向作者/读者索取更多资源

PurposeNeutrons are an unavoidable by-product of high-energy radiation therapy treatments that deliver unwanted nontarget dose to patients. Use of flattening-filter-free (FFF) photon beams has been shown to significantly reduce photoneutron production per monitor unit (MU) of dose delivered. The purpose of this investigation was to characterize the photoneutron production of the 10MV and 10MV FFF beams of the Varian TrueBeam(TM) linear accelerator. MethodsNeutron fluence spectra were measured using a Nested Neutron Spectrometer(TM) (NNS, Detec Inc., Gatineau, Canada). The ratios of neutron fluence and ambient dose equivalent for the 10MV FFF beam relative to the 10MV beam, dubbed FF-ratios (FFF/FF), were used to characterize the difference between the two beams. FF-ratios were compared under the following three conditions (a) per MU, at various locations in the treatment room, (b) per MU, with the linac jaws opened and closed, and (c) per electron striking the bremsstrahlung target, as opposed to per MU, at one location with the jaws closed. ResultsOn average, the neutron fluence for the 10MV FFF beam was 37% lower per MU than the 10MV beam (FF-ratio=0.63). The FF-ratio in neutron fluence and ambient dose equivalent did not vary by much between different locations within the treatment room. However, the FF-ratio in neutron ambient dose equivalent was reduced significantly when the linac jaws were opened compared to closed, which implies that the jaws contribute more to the photoneutron spectrum of the 10MV FFF beam than to the 10MV beam. Finally, it was found that the 10MV FFF beam produces more photoneutrons per electron striking the bremsstrahlung target than the 10MV beam (FF-ratio=2.56). ConclusionsThe photoneutron fluence per MU produced by the 10MV FFF beam is 37% lower than the 10MV beam of a Varian TrueBeam linac. Accordingly, a reduction in neutron dose received by patients is achieved through use of the unflattened beam, provided that treatment plans for each beam require approximately the same number of MU. It was found to be instructive to compare the photoneutron yield per source electron between the two beams as it helped provide an understanding of the physics underlying photoneutron production in both beams.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据