Low-dose phase contrast tomography with conventional x-ray sources

期刊

MEDICAL PHYSICS
卷 41, 期 7, 页码 -

出版社

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.4884297

关键词

tomography; x-rays; phase contrast

资金

  1. UK Engineering and Physical Sciences Research Council [EP/L001381/1, EP/I021884/1]
  2. Marie Curie Career Integration Grant within the Seventh Framework Programme of the European Union [PCIG12-GA-2012-333990, PCIG12-GA-2012-334056]
  3. Discovery Early Career Research Award from the Australian Research Council [DE120101331]
  4. EPSRC [EP/I021884/1, EP/L001381/1] Funding Source: UKRI
  5. Engineering and Physical Sciences Research Council [EP/I021884/1, EP/L001381/1] Funding Source: researchfish

向作者/读者索取更多资源

Purpose: The edge illumination (EI) x-ray phase contrast imaging (XPCi) method has been recently further developed to perform tomographic and, thus, volumetric imaging. In this paper, the first tomographic EI XPCi images acquired with a conventional x-ray source at dose levels below that used for preclinical small animal imaging are presented. Methods: Two test objects, a biological sample and a custom-built phantom, were imaged with a laboratory-based EI XPCi setup in tomography mode. Tomographic maps that show the phase shift and attenuating properties of the object were reconstructed, and analyzed in terms of signal-to-noise ratio and quantitative accuracy. Dose measurements using thermoluminescence devices were performed. Results: The obtained images demonstrate that phase based imaging methods can provide superior results compared to attenuation based modalities for weakly attenuating samples also in 3D. Moreover, and, most importantly, they demonstrate the feasibility of low-dose imaging. In addition, the experimental results can be considered quantitative within the constraints imposed by polychromaticity. Conclusions: The results, together with the method's dose efficiency and compatibility with conventional x-ray sources, indicate that tomographic EI XPCi can become an important tool for the routine imaging of biomedical samples. (C) 2014 American Association of Physicists in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据