Validation of simplified dosimetry approaches in 89 Zr-PET/CT: The use of manual versus semi-automatic delineation methods to estimate organ absorbed doses

期刊

MEDICAL PHYSICS
卷 41, 期 10, 页码 -

出版社

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.4895973

关键词

PET/CT; Zr-89; image registration; dosimetry; absorbed dose

资金

  1. Philips Healthcare

向作者/读者索取更多资源

Purpose: Increasing interest in immuno-positron emission tomography (PET) studies requires development of dosimetry methods which will provide accurate estimations of organ absorbed doses. The purpose of this study is to develop and validate simplified dosimetry approaches for (89)Zirconium-PET (Zr-PET)/computed tomography (CT) studies. Methods: Five patients with advanced colorectal cancer received 37.1 +/- 0.9 MBq Zr-89-cetuximab within 2 h after administration of a therapeutic dose of 500 mg m(-2) cetuximab. PET/CT scans were obtained 1, 24, 48, 94, and 144 h post injection. Volumes of interest (VOIs) were manually delineated in lungs, liver, spleen, and kidneys for all scans, providing a reference VOI set. Simplified manual VOIs were drawn independently on CT scans using larger voxel sizes. The transformation of VOIs based on rigid and/or nonrigid registrations of the first CT scan (CT1) onto all successive CT scans was also investigated. The transformation matrix obtained from each registration was applied to the manual VOIs of CT1 to obtain VOIs for the successive scans. Dice similarity coefficient (DSC) and Hausdorff distance were used to assess the performance of the registrations. Organ total activity, organ absorbed dose, and effective dose were calculated for all methods. Results: Semi-automatic delineation based on nonrigid registration showed excellent agreement for lungs and liver (DSC: 0.90 +/- 0.04; 0.81 +/- 0.06) and good agreement for spleen and kidneys (DSC: 0.71 +/- 0.07; 0.66 +/- 0.08). Hausdorff distance ranged from 13 to 16 mm depending on the organ. Simplified manual delineation methods, in liver and lungs, performed similarly to semi-automatic delineation methods. For kidneys and spleen, however, poorer accuracy in total activity and absorbed dose was observed, as the voxel size increased. Organ absorbed dose and total activity based on nonrigid registration were within 10%. The effective dose was within +/- 3% for all VOI delineation methods. Conclusions: A fast, semi-automatic, and accurate delineation method based on nonrigid registration was developed for determination of organ absorbed and effective dose in Zr-89-PET/CT studies which may also be applied to other long-lived radionuclide PET/CT studies. (C) 2014 American Association of Physicists in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据