Evaluation of motion correction methods in human brain PET imaging-A simulation study based on human motion data

期刊

MEDICAL PHYSICS
卷 40, 期 10, 页码 -

出版社

WILEY
DOI: 10.1118/1.4819820

关键词

PET; motion correction; event-by-event; dynamic; frame-based

资金

  1. National Institute of Neurological Disorders and Stroke [R01NS058360]
  2. National Center for Research Resources (NCRR) [UL1 RR024139]
  3. National Center for Advancing Translational Science (NCATS), components of the National Institutes of Health (NIH)
  4. NIH roadmap for Medical Research

向作者/读者索取更多资源

Purpose: Motion correction in PET has become more important as system resolution has improved. The purpose of this study was to evaluate the accuracy of event-by-event and frame-based MC methods in human brain PET imaging. Methods: Motion compensated image reconstructions were performed with static and dynamic simulated high resolution research tomograph data with frame-based image reconstructions, using a range of measured human head motion data. Image intensities in high-contrast regions of interest (ROI) and parameter estimates in tracer kinetic models were assessed to evaluate the accuracy of the motion correction methods. Results: Given accurate motion data, event-by-event motion correction can reliably correct for head motions. The average ROI intensities and the kinetic parameter estimates V-T and BPND were comparable to the true values. The frame-based motion correction methods with correctly aligned attenuation map using the average of externally acquired motion data or motion data derived from image registration give comparable quantitative accuracy. For large intraframe (>5 mm) motion, the frame-based methods produced similar to 9% bias in ROI intensities, similar to 5% in VT, and similar to 10% in BPND estimates. In addition, in real studies that lack a ground truth, the normalized weighted residual sum of squared difference is a potential figure-of-merit to evaluate the accuracy of motion correction methods. Conclusions: The authors conclude that frame-based motion correction methods are accurate when the intraframe motion is less than 5 mm and when the attenuation map is accurately aligned. Given accurate motion data, event-by-event motion correction can reliably correct for head motion in human brain PET studies. (C) 2013 American Association of Physicists in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据