4.6 Article

A variable echo-number method for estimating R2 in MRI-based polymer gel dosimetry

期刊

MEDICAL PHYSICS
卷 38, 期 2, 页码 975-982

出版社

WILEY
DOI: 10.1118/1.3544659

关键词

quantitative MRI; CPMG; polymer gel dosimetry; maximum likelihood estimator; R2; radiation dose

资金

  1. Minnesota Medical Foundation
  2. NIH [P30 NS057091, P41 RR008079]

向作者/读者索取更多资源

Purpose: Spin-spin relaxation rate R2 is commonly used to quantify absorbed dose for magnetic resonance imaging (MRI)-based polymer gel dosimetry. R2 is estimated by applying a parameter fitting algorithm to a train of spin-echo signals. However, a careless application of a large number of echoes can result in anomalous R2 values because the echo signal intensity decreases to the background signal offset level for a long echo time. In this article, the authors proposed and evaluated a variable echo-number (VAREC) method to remedy the problem. Methods: The VAREC algorithm uses only echo signals, whose intensities are greater than a preset threshold. Here, the threshold is defined as the standard deviation of Gaussian noise times a multiplier alpha. The authors implemented three R2 estimation methods in an in-house program: The nonlinear least-squares algorithm (NLLS), the VAREC method, and the maximum likelihood estimator with the Rician signal intensity distribution (MLE_R). Those methods were used to estimate the R2 values of test phantoms with known R2 values and BANG3-type polymer gels, which were irradiated to 12 different doses ranging from 0 to 50 Gy. The R2 values were measured by using a 32-echo CPMG pulse sequence on 3 T MRI scanners. The R2 values of the VAREC method were compared with those of NLLS and MLE_R. Results: The R2 values of the NLLS method incorrectly decreased to the zero-dose level for doses greater than 10 Gy. The R2 values of the VAREC method with alpha = 2 agreed with those of MLE_R within the measurement uncertainty. The uncertainties of the R2 values were the smallest for alpha = 2 or 3 among various alpha values. Conclusions: The VAREC algorithm is simple, fast, and robust for the R2 estimation. The authors recommend this method with alpha = 2 or 3 for R2 estimation using multispin echo MRI protocols. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3544659]

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据