4.6 Article

Lung dosimetry in a linac-MRI radiotherapy unit with a longitudinal magnetic field

期刊

MEDICAL PHYSICS
卷 37, 期 9, 页码 4722-4732

出版社

WILEY
DOI: 10.1118/1.3475942

关键词

Linac-MR; MRI-radiotherapy; patient dosimetry; magnetic field; longitudinal; transverse; MRI-guided radiotherapy; Monte Carlo

资金

  1. Alberta Cancer Foundation
  2. Alberta Health Services

向作者/读者索取更多资源

Purpose: There is interest in developing linac-MR systems for MRI-guided radiation therapy. To date, the designs for such linac-MR devices have been restricted to a transverse geometry where the static magnetic field is oriented perpendicular to the direction of the incident photon beam. This work extends possibilities in this field by proposing and examining by Monte Carlo simulations, a probable longitudinal configuration where the magnetic field is oriented in the same direction as the photon beam. Methods: The EGSnrc Monte Carlo (MC) radiation transport codes with algorithms implemented to account for the magnetic field deflection of charged particles were used to compare dose distributions for linac-MR systems in transverse and longitudinal geometries. Specifically, the responses to a 6 MV pencil photon beam incident on water and lung slabs were investigated for 1.5 and 3.0 T magnetic fields. Further a five field lung plan was simulated in the longitudinal and transverse geometries across a range of magnetic field strengths from 0.2 through 3.0 T. Results: In a longitudinal geometry, the magnetic field is shown to restrict the radial spread of secondary electrons to a small degree in water, but significantly in low density tissues such as lung in contrast to the lateral shift in dose distribution seen in the transverse geometry. These effects extend to the patient case, where the longitudinal configuration demonstrated dose distributions more tightly confined to the primary photon fields, which increased dose to the planning target volume (PTV), bettered dose homogeneity within a heterogeneous (in density) PTV, and reduced the tissue interface effects associated with the transverse geometry. Conclusions: Dosimetry issues observed in a transverse linac-MR geometry such as changes to the depth dose distribution and tissue interface effects were significantly reduced or eliminated in a longitudinal geometry on a representative lung plan. Further, an increase in dose to the PTV, resulting from the magnetic field confining electrons to the forward direction, shows potential for a reduction in dose to the surrounding tissues. (c) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3475942]

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据