Unintended attenuation in the Leksell Gamma Knife® Perfexion™ calibration-phantom adaptor and its effect on dose calibration

期刊

MEDICAL PHYSICS
卷 36, 期 4, 页码 1208-1211

出版社

AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.3093240

关键词

calibration; dosimeters; dosimetry; phantoms

向作者/读者索取更多资源

The calibration of Leksell Gamma Knife Perfexion (LGK PFX) is performed using a spherical polystyrene phantom 160 mm in diameter, which is provided by the manufacturer. This is the same phantom that has been used with LGK models U, B, C, and 4C. The polystyrene phantom is held in irradiation position by an aluminum adaptor, which has stainless steel side-fixation screws. The phantom adaptor partially attenuates the beams from sectors 3 and 7 by 3.2% and 4.6%, respectively. This unintended attenuation introduces a systematic error in dose calibration. The overall effect of phantom-adaptor attenuation on output calibration of the LGK PFX unit is to underestimate output by about 1.0%.