MEDICAL PHYSICS
卷 35, 期 4, 页码 1494-1496出版社
AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.2889620
关键词
CyberKnife; radiosurgery; peripheral dose
The authors investigated the peripheral dose reduction for CyberKnife radiosurgery treatments after the installation of a linac shielding upgrade. As in a previous investigation, the authors considered two treatment plans, one for a hypothetical target in the brain and another for a target in the thorax, delivered to an anthropomorphic phantom. The results of the prior investigation showed that the CyberKnife delivered significantly higher peripheral doses than comparable model C Gamma Knife or IMRT treatments. Current measurements, after the linac shielding upgrade, demonstrate that the additional shielding decreased the peripheral dose, expressed as a percentage of the delivered monitor units (MU), by a maximum of 59%. The dose reduction was greatest for cranial-caudal distances from the field edge less than 30 cm, and at these distances, the CyberKnife peripheral dose, expressed as a percentage of the delivered MU, is now comparable to that measured for the other treatment modalities in our previous investigation. For distances between 30 and 70 cm from the field edge, the additional shielding reduced the peripheral dose by between 20% and 55%. At these distances, the CyberKnife peripheral dose remains higher than doses measured in our previous study for the model C Gamma Knife and IMRT. (c) 2008 American Association of Physicists in Medicine.