4.6 Article Proceedings Paper

Clinical prototype of a plastic water-equivalent scintillating fiber dosimeter array for QA applications

期刊

MEDICAL PHYSICS
卷 35, 期 8, 页码 3682-3690

出版社

WILEY
DOI: 10.1118/1.2953564

关键词

plastic scintillators; dosimeter array; dosimetry; IMRT

向作者/读者索取更多资源

A clinical prototype of a scintillating fiber dosimeter array for quality assurance applications is presented. The array consists of a linear array of 29 plastic scintillation detectors embedded in a water-equivalent plastic sheet coupled to optical fibers used to guide optical photons to a charge coupled device (CCD) camera. The CCD is packaged in a light-tight, radiation-shielded housing designed for convenient transport. A custom designed connector is used to ensure reproducible mechanical positioning of the optical fibers relative to the CCD. Profile and depth dose characterization measurements are presented and show that the prototype provides excellent dose measurement reproducibility (+/- 0.8%) in-field and good accuracy (+/- 1.6% maximum deviation) relative to the dose measured with an IC10 ionization chamber.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据