4.4 Article

A numerical investigation into the influence of the properties of cobalt chrome cellular structures on the load transfer to the periprosthetic femur following total hip arthroplasty

期刊

MEDICAL ENGINEERING & PHYSICS
卷 36, 期 4, 页码 458-466

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.medengphy.2014.02.008

关键词

Femoral stem; Cellular structures; Finite element analysis

向作者/读者索取更多资源

Stress shielding of the periprosthetic femur following total hip arthroplasty is a problem that can promote the premature loosening of femoral stems. In order to reduce the need for revision surgery it is thought that more flexible implant designs need to be considered. In this work, the mechanical properties of laser melted square pore cobalt chrome molybdenum cellular structures have been incorporated into the design of a traditional monoblock femoral stem. The influence of incorporating the properties of cellular structures on the load transfer to the periprosthetic femur was investigated using a three dimensional finite element model. Eleven different stiffness configurations were investigated by using fully porous and functionally graded approaches. This investigation confirms that the periprosthetic stress values depend on the stiffness configuration of the stem. The numerical results showed that stress shielding is reduced in the periprosthetic Gruen zones when the mechanical properties of cobalt chrome molybdenum cellular structures are used. This work identifies that monoblock femoral stems manufactured using a laser melting process, which are designed for reduced stiffness, have the potential to contribute towards reducing stress shielding. (C) 2014 IPEM. Published by Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据