期刊
MEDICAL ENGINEERING & PHYSICS
卷 33, 期 3, 页码 281-289出版社
ELSEVIER SCI LTD
DOI: 10.1016/j.medengphy.2010.10.011
关键词
Finite element analysis (FEA); Patient-specific model; Carotid artery stenting (CAS)
资金
- MIUR (Ministero dell'Istruzione, dell'Universita e della Ricerca, Italian Ministry of Education, University and Research)
The outcome of carotid artery stenting (CAS) depends on a proper selection of patients and devices, requiring dedicated tools able to relate the device features with the target vessel. In the present study, we use finite element analysis to evaluate the performance of three self-expanding stent designs (laser-cut open-cell, laser-cut closed-cell, braided closed-cell) in a carotid artery (CA). We define six stent models considering the three designs in different sizes and configurations (i.e. straight and tapered), evaluating the stress induced in the vessel wall, the lumen gain and the vessel straightening in a patient-specific CA model based on computed angiography tomography (CTA) images. For the considered vascular anatomy and stents, the results suggest that: (i) the laser-cut closed-cell design provides a higher lumen gain; (ii) the impact of the stent configuration and of the stent oversizing is negligible with respect to the lumen gain and relevant with respect to the stress induced in the vessel wall; (iii) stent design, configuration and size have a limited impact on the vessel straightening. The presented numerical model represents a first step towards a quantitative assessment of the relation between a given carotid stent design and a given patient-specific CA anatomy. (C) 2010 IPEM. Published by Elsevier Ltd. All rights reserved.