4.5 Article

Energy harvesting through arterial wall deformation: design considerations for a magneto-hydrodynamic generator

期刊

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s11517-012-0989-2

关键词

Human energy scavenging; Electromagnetism; Fluid-structure interactions; Arterial pulse propagation; Analytical model

向作者/读者索取更多资源

As the complexity of active medical implants increases, the task of embedding a life-long power supply at the time of implantation becomes more challenging. A periodic renewal of the energy source is often required. Human energy harvesting is, therefore, seen as a possible remedy. In this paper, we present a novel idea to harvest energy from the pressure-driven deformation of an artery by the principle of magneto-hydrodynamics. The generator relies on a highly electrically conductive fluid accelerated perpendicularly to a magnetic field by means of an efficient lever arm mechanism. An artery with 10 mm inner diameter is chosen as a potential implantation site and its ability to drive the generator is established. Three analytical models are proposed to investigate the relevant design parameters and to determine the existence of an optimal configuration. The predicted output power reaches 65 mu W according to the first two models and 135 mu W according to the third model. It is found that the generator, designed as a circular structure encompassing the artery, should not exceed a total volume of 3 cm(3).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。