4.5 Article

Platinum electrode noise in the ENG spectrum

期刊

MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING
卷 46, 期 10, 页码 997-1003

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s11517-008-0386-z

关键词

electrode noise; Pt electrode; impedance measurement; implantable neuroprostheses; neural signal recording; noise measurement

资金

  1. UK Engineering and Physical Sciences Research Council [GR/S93790/01]
  2. European Union
  3. Engineering and Physical Sciences Research Council [EP/F009593/1, GR/S93790/01] Funding Source: researchfish
  4. EPSRC [EP/F009593/1] Funding Source: UKRI

向作者/读者索取更多资源

The noise from Pt electrodes has not been measured before. We measured the impedance and noise between pairs of Pt electrodes in cylindrical cells within frequency bands which contain the electroneurogram band (ENG, 500-10 kHz). The results show that the noise from the Pt-saline interface and also the volume conductor are the same as the thermal noise from the real part of those impedances. This means that Pt electrodes are convenient for neural recording because the total source noise from the electrodes and the volume conductor (tissue) can be estimated from the measured total impedance and calculating the thermal noise associated with the resistive part. This differs from Ag/AgCl electrodes for which the interface noise has been reported to be higher than the thermal noise, and therefore calculation of the total noise will be more difficult.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据