4.5 Article

Toward calibration-free wall shear stress measurement using a dual hot-film sensor and Kelvin bridges

期刊

MEASUREMENT SCIENCE AND TECHNOLOGY
卷 29, 期 10, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1361-6501/aadb1b

关键词

hot-film; MEMS thermal sensor; wall shear stress; skin friction; calibration-free

资金

  1. National Natural Science Foundation of China [11572078, 91752101]
  2. 973 Plan [2014CB744100]

向作者/读者索取更多资源

Combined heat transfer and wall shear stress measurements on the inner wall of a fully developed turbulent air flow in a round pipe were presented in this paper. The heat transfer was measured using a flush-mounted dual hot-film sensor composed of a non-electric conductive membrane sandwiched in between two thin nickel films. Each of the two films was installed in a branch of a separate Kelvin bridge, and operated at a same temperature; the bottom film served as an active heat insulator so that the Joule heat from the upper film transferred only to the air. Both theoretical analysis and measurements indicated (Nu) over bar . Pe (-1/3) approximate to 0.91 for 300 <= Pe <= 2500, where (Nu) over bar was the averaged Nusselt number and Pe was the Peclet number. A calibration-free technique for wall shear stress measurement based only on measuring the Joule heat flux was found to be feasible, providing the Peclet number was in the range of 300-2500 and the film temperature was sufficiently large.