4.5 Article

An acetone breath analyzer using cavity ringdown spectroscopy: an initial test with human subjects under various situations

期刊

MEASUREMENT SCIENCE AND TECHNOLOGY
卷 19, 期 10, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0957-0233/19/10/105604

关键词

breath analyzer; CRDS; background subtraction method; acetone; diabetes; alcohol

资金

  1. National Science Foundation [CTS-0626302]

向作者/读者索取更多资源

We have developed a portable breath acetone analyzer using cavity ringdown spectroscopy (CRDS). The instrument was initially tested by measuring the absorbance of breath gases at a single wavelength (266 nm) from 32 human subjects under various conditions. A background subtraction method, implemented to obtain absorbance differences, from which an upper limit of breath acetone concentration was obtained, is described. The upper limits of breath acetone concentration in the four Type 1 diabetes (T1D) subjects, tested after a 14 h overnight fast, range from 0.80 to 3.97 parts per million by volume (ppmv), higher than the mean acetone concentration (0.49 ppmv) in non-diabetic healthy breath reported in the literature. The preliminary results show that the instrument can tell distinctive differences between the breath from individuals who are healthy and those with T1D. On-line monitoring of breath gases in healthy people post-exercise, post-meals and post-alcohol-consumption was also conducted. This exploratory study demonstrates the first CRDS-based acetone breath analyzer and its potential application for point-of-care, non-invasive, diabetic monitoring.