MACS - a new high intensity cold neutron spectrometer at NIST

期刊

MEASUREMENT SCIENCE AND TECHNOLOGY
卷 19, 期 3, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0957-0233/19/3/034023

关键词

neutron instrumentation; neutron diffraction; neutron spectrometer; neutron scattering

向作者/读者索取更多资源

We describe a novel cold neutron spectrometer under development at NIST optimized for wave vector resolved spectroscopy with incident energies between 2.1 meV and 20 meV and energy resolution from 0.05 meV (E-i = 2.1 meV) to 3.0 meV (E-i = 20 meV). By using a 1428 cm(2) double focusing PG (0 0 2) monochromator close to the National Institute of Standards and Technology (NIST) cold neutron source the instrument provides up to 5 x 10(8) neutrons cm(-2) s(-1) on a 8 cm(2) sample area. The measured performance is consistent with Monte Carlo simulations. The monochromating system, which includes radial collimators, three filters and a variable beam aperture, offers considerable flexibility in optimizing Q-resolution, energy resolution and intensity. The detector system will consist of an array of 20 channels which combined will subtend a solid angle of 0.2 sr. This is approximately a factor of 40 more than a conventional triple axis spectrometer. Each detector channel contains a vertically focusing double crystal analyzer system (DXAL) actuated by a single stepping motor. We find identical integrated reflectivity at approximately 10% coarser energy resolution for the 130' mosaic double bounce analyzer as compared to a conventional 25' analyzer at the same energy. The vertical focusing of the DXAL allows for smaller detectors for enhanced signal to noise with 8 degrees. vertical acceptance. Options for post sample collimators and filters provide flexibility in the choice of scattered beam energy and wavevector resolution.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据