期刊
MEASUREMENT
卷 45, 期 6, 页码 1670-1673出版社
ELSEVIER SCI LTD
DOI: 10.1016/j.measurement.2012.03.020
关键词
Standard addition; Calibration; Precision; Extrapolation; Comparison
资金
- UK Department of Business, Innovation and Skills
Two quantification strategies for one-point standard addition calibration have been compared mathematically. One strategy involved the extrapolation of measurement points to their intercept with the x-axis to determine the analyte content in the unknown sample, and the other strategy is based upon direct calculation of the analyte content in the unknown sample using the instrumental responses obtained during measurement. The cases of both conventional standard addition calibration (C-SAC) and sequential standard addition calibration (S-SAC) have been considered. The heteroscedastic situation has been considered, where the relative precision of instrumental responses is constant. (C) 2012 Elsevier Ltd. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐