4.5 Article

Development of a vector-tensor system to measure the absolute magnetic flux density and its gradient in magnetically shielded rooms

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 86, 期 5, 页码 -

出版社

AIP Publishing
DOI: 10.1063/1.4921583

关键词

-

资金

  1. DFG Priority Program [SPP 1491, BU-1037/3-2]

向作者/读者索取更多资源

Several experiments in fundamental physics demand an environment of very low, homogeneous, and stable magnetic fields. For the magnetic characterization of such environments, we present a portable SQUID system that measures the absolute magnetic flux density vector and the gradient tensor. This vector-tensor system contains 13 integrated low-critical temperature (LTc) superconducting quantum interference devices (SQUIDs) inside a small cylindrical liquid helium Dewar with a height of 31 cm and 37 cm in diameter. The achievable resolution depends on the flux density of the field under investigation and its temporal drift. Inside a seven-layer mu-metal shield, an accuracy better than +/- 23 pT for the components of the static magnetic field vector and +/- 2 pT/cm for each of the nine components of the gradient tensor is reached by using the shifting method. (C) 2015 AIP Publishing LLC.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据