4.5 Article

Blood triggered corrosion of magnesium alloys

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.mseb.2011.06.006

关键词

Stent; Mg release; Chandler-Loop; Static immersion test; Blood corrosion; Simulated body fluid (SBF)

资金

  1. AiF [KF0548101PK7]

向作者/读者索取更多资源

Intravascular stents manufactured out of bioabsorbable magnesium (Mg) or Mg-alloys are considered as auspicious candidates for the next stent generation. However, before clinical application numerous physical and biological tests, especially to predict the clinically highly important degradation kinetics in vivo, have to be performed. In a Chandler-Loop model, the initial degradation of eight different magnesium alloys during 6h in contact with human whole blood was investigated. The magnesium release varied between 0.91 +/- 0.33 mg/cm(2) (MgAl9Zn1) and 2.57 0.38 mg/cm(2) (MgZn1). No correlation could be found with Mg release data obtained after immersion in simulated body fluid (SBF). This pilot study showed that Mg corrosion is highly influenced by the biological test environment (SBF or blood, etc.) and that a modified Chandler-Loop model with human whole blood may be superior to predict corrosion of Mg alloys under clinical conditions than the SBF models presently used. (C) 2011 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据