4.2 Article

In vivo measurement of glycine with short echo-time 1H MRS in human brain at 7 T

出版社

SPRINGER
DOI: 10.1007/s10334-008-0152-0

关键词

Glycine; Magnetic resonance spectroscopy; 7T; Human brain

资金

  1. Centre d'Imagerie BioMedicale (CIBM)
  2. UNIL
  3. UNIGE
  4. HUG
  5. CHUV
  6. EPFL
  7. Leenaards and Jeantet Foundations
  8. Swiss National Science Foundation [3100A0-116220]

向作者/读者索取更多资源

To determine whether glycine can be measured at 7 T in human brain with H-1 magnetic resonance spectroscopy (MRS). The glycine singlet is overlapped by the larger signal of myo-inositol. Density matrix simulations were performed to determine the TE at which the myo-inositol signal was reduced the most, following a single spin-echo excitation. H-1 MRS was performed on an actively shielded 7 T scanner, in five healthy volunteers. At the TE of 30 ms, the myo-inositol signal intensity was substantially reduced. Quantification using LCModel yielded a glycine-to-creatine ratio of 0.14 +/- A 0.01, with a Cram,r-Rao lower bound (CRLB) of 7 +/- 1%. Furthermore, quantification of metabolites other than glycine was possible as well, with a CRLB mostly below 10%. It is possible to detect glycine at 7 T in human brain, at the short TE of 30 ms with a single spin-echo excitation scheme.