The Effect of Low b-Values on the Intravoxel Incoherent Motion Derived Pseudodiffusion Parameter in Liver

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 73, 期 1, 页码 306-311

出版社

WILEY-BLACKWELL
DOI: 10.1002/mrm.25109

关键词

b-value; pseudodiffusion; liver; diffusion

资金

  1. Advancing a Healthier Wisconsin [3R01 CA082500]

向作者/读者索取更多资源

PurposeTo examine the effect of low b-values (0<50 s/mm(2)) on the calculation of the intravoxel incoherent motion (IVIM) derived pseudodiffusion parameter in the normal liver. MethodsSimulations were performed to examine the effects of adding low b-values on the pseudodiffusion parameter. Low b-values were cumulatively added to the distribution and the IVIM signal was generated with varying pseudodiffusion values. The signal was fit with the IVIM model after the addition of Gaussian noise, and the simulated values were compared with the true values. In addition, the livers of eight control subjects were imaged using respiratory-triggered DWI. Pseudodiffusion was calculated with and without low b-values and compared. ResultsPseudodiffusion tended to be underestimated when low b-values were not included in the b-value distribution as predicted by simulations and confirmed with in vivo imaging. The number of outlier values was also reduced as more low b-values were added. ConclusionIn conclusion, this study showed pseudodiffusion in the liver tended to be underestimated when too few low b-values (0<50 s/mm(2)) were included in the distribution. Therefore, it is recommended to include at least two low b-values when performing liver IVIM studies. Magn Reson Med 73:306-311, 2015. (c) 2014 Wiley Periodicals, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据