4.5 Article

Volumetric measurement of perfusion and arterial transit delay using hadamard encoded continuous arterial spin labeling

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 69, 期 4, 页码 1014-1022

出版社

WILEY
DOI: 10.1002/mrm.24335

关键词

-

资金

  1. National Institues of Health [T32AG023480, MH80729, CA115745]

向作者/读者索取更多资源

Creating images of the transit delay from the labeling location to image tissue can aid the optimization and quantification of arterial spin labeling perfusion measurements and may provide diagnostic information independent of perfusion. Unfortunately, measuring transit delay requires acquiring a series of images with different labeling timing that adds to the time cost and increases the noise of the arterial spin labeling study. Here, we implement and evaluate a proposed Hadamard encoding of labeling that speeds the imaging and improves the signal-to-noise ratio efficiency. Volumetric images in human volunteers confirmed the theoretical advantages of Hadamard encoding over sequential acquisition of images with multiple labeling timing. Perfusion images calculated from Hadamard encoded acquisition had reduced signal-to-noise ratio relative to a dedicated perfusion acquisition with either assumed or separately measured transit delays, however. Magn Reson Med 69:10141022, 2013. (c) 2012 Wiley Periodicals, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据