MR safety: Fast T1 thermometry of the RF-induced heating of medical devices

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 68, 期 5, 页码 1593-1599

出版社

WILEY-BLACKWELL
DOI: 10.1002/mrm.24171

关键词

MR thermometry; inversion recovery snapshot FLASH; T1 temperature dependency; sliding-window

资金

  1. Bavarian Research Foundation [BFS:AZ-828-08]

向作者/读者索取更多资源

Determining the MR compatibility of medical implants and devices is becoming increasingly relevant. In most cases, the heating of conductive implants due to radiefrequency (RF) excitation pulses is measured by fluoroptic temperature sensors in relevant tests for approval. Another common method to determine these heating effects is MR thermometry using the proton resonance frequency. This method gives good results in homogeneous phantoms. However in many cases, technical shortcomings such as susceptibility artifacts prohibit exact proton resonance frequency thermometry near medical implants. Therefore, this work aimed at developing a fast T1-based method which allows controlled MR-related heating of a medical implant while simultaneously quantifying the spatial and temporal temperature distribution. To this end, an inversion recovery snapshot Fast Low-Angle Shot (FLASH) sequence was modified with additional off-resonant heating pulses. With an accelerated imaging method and a sliding-window technique, every 7.6 s a new temperature map could be generated with a spatial in-plane resolution of 2 mm. The temperature deviation from calculated temperature values to reference fluoroptic probe was found to be smaller than 1 K. Magn Reson Med, 2012. (c) 2012 Wiley Periodicals, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据